Operator’s Manual – OPTI CCA-TS 3-15
3 CUSTOMIZATION
3.3.2.7 DeningUnits
This menu lets you change the units of measure
for pressure, temperature, output resolution, total
hemoglobin, Ca
++
, Glu, BUN (urea) and Lac.
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press <Units>
(Fig. 3-30).
3. In the <Units> screen, select the units for the
displayed parameters (Fig. 3-31).
4. The selection for <Resolution> on this menu
determines the number of digits displayed and
printed past the decimal point, for all measured
parameters.
NOTE: The selection applies to sample results only.
Resolution is always High for Control and
SRC results. Resolution examples are shown
in the following table:
Low
High
pH 7.34 pH 7.341
PCO
2
43 mmHg PCO
2
43.2 mmHg
PO
2
87 mmHg PO
2
86.8 mmHg
Na
+
143 mmol/L Na
+
143.3 mmol/L
K
+
4.6 mmol/L K
+
4.57 mmol/L
Cl
-
103 mmol/L Cl
-
103.1 mmol/L
Ca
++
1.21 mmol/L Ca
++
1.21 mmol/L
Glu 5.71 mmol/L Glu 5.71 mmol/L
BUN 18.5 mg/dL BUN 18.5 mg/dL
Lac 14.5 mmol/L Lac 14.5 mmol/L
tHb 14.6 g/dL tHb 14.6 g/dL
SO
2
90 % SO
2
89.8 %
NOTE: PO
2
and PCO
2
values above 100 mmHg
are always displayed to the nearest whole
number.
Fig. 3-30 Select Units
Fig. 3-31 Select Units