Operator’s Manual – OPTI CCA-TS 4-7
4 CALIBRATION AND QUALITY CONTROL
4.5.2 Printing SRC Results
This menu allows you to print out SRC reports or
SRC statistical information.
1. From the <Ready> display, select
<DataManager>.
2. Select <SRC> (Fig. 4-10).
3. In the <Data-SRCMeasurement> screen
(Fig. 4-11), press the
button to display
the SRC results (Fig. 4-12). Use the
and buttons to display the previous or
next page of results.
4. To print individual results, highlight the desired
measurement (Fig. 4-11). To print groups
of results, highlight the rst measurement
to be printed, press
, then select
the last measurement to be printed. All the
measurements in between will be selected.
Press
to select all results.
5. Press to print your selection.
6. Press the <Statistics> button to print out the
statistics from the most recent 30 days of SRC
measurements for all levels.
7. After printout, the database can be deleted by
pressing .
8. If a password has been activated under
<Setup>, you must enter it at this time before
the data is deleted.
The unit will now delete all SRC data from the
internal database.
9. Press to return to the <Ready> display.
Fig. 4-10 Select SRC
Fig. 4-11 SRC Measurements
Fig. 4-12 View SRC Results

Table of Contents