4-6 Operator’s Manual – OPTI CCA-TS
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-7 SRC Measurement
After you have veried that the SRC information
is correct, the instrument begins the measurement
process which is indicated on the display screen
(Fig. 4-7). During this time (about 60 seconds),
a progress bar is displayed.
7. When the measurement is complete, the unit
displays the results (Fig. 4-8 and Fig. 4-9).
NOTE: Resultsobtainedareapplicabletothesensor
cassettetypebeingusedforpatientsample.
• The unit automatically checks the results
against the ranges and stores the results in its
internal database.
• For parameters within range, <Pass> will
be displayed and printed.
• For parameters out of range, or if an internal
drift is detected, <Fail> will be displayed.
8. Open the sample chamber cover and remove the
SRC.
9. Place the SRC back into its pouch immediately
after removal from the instrument.
10. Close the sample chamber cover.
• If the SRC test failed, gently clean the SRC,
the optics window, and the inside cover of
the SMC with alcohol and a lint-free cloth
and repeat this process. If it fails again,
refer to the troubleshooting section of this
manual.
• Perform the second SRC measurement with
another SRC Level in the same manner. If
both SRC tests passed, the unit is ready to
perform measurements.
NOTE: ForapplicationofQCLockout,pleaserefer
tosection3.3.3.1.2.
NOTE: Verifywithyourparticularregulatory
agencyandyourinternalpolicyregarding
numberoflevelsandfrequencyofSRCs
toberun.Athirdlevel(normalrange)of
SRCisavailableasanoption(PartNumber
BP7554).