EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - Checking the Optics

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7-20 Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
Fig. 7-18 Optics Test
Fig. 7-19 Remove SRC
7.2.13 Checking the Optics
Fig. 7-16 Optics Test
Fig. 7-17 Insert SRC
This option checks the output of the six optics
channels. This test is designed for trained service
personnel.
From the <Ready> display, press <System
Manager> and <Diagnostics>.
• Select the <Tests> tab.
Press <Optics> to start the test (Fig. 7-16).
• Insert an SRC and press
(Fig. 7-17).
• The system will now check the optics
(Fig. 7-18).
• At the completion of the test, a printout of the
results will be printed and the <Remove SRC>
message will be displayed (Fig. 7-19).
Test in Process . . .
Optics Test . . .
Test in Process . . .
Optics Test . . .
Remove SRC
Optics Test . . .
Remove SRC
Optics Test . . .
Ins ert SRC
Optics Test . . .
OK
Cancel
Ins ert SRC
Optics Test . . .
OK
Cancel

Table of Contents