7-20 Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
Fig. 7-18 Optics Test
Fig. 7-19 Remove SRC
7.2.13 Checking the Optics
Fig. 7-16 Optics Test
Fig. 7-17 Insert SRC
This option checks the output of the six optics
channels. This test is designed for trained service
personnel.
From the <Ready> display, press <System
Manager> and <Diagnostics>.
• Select the <Tests> tab.
•
Press <Optics> to start the test (Fig. 7-16).
• Insert an SRC and press
(Fig. 7-17).
• The system will now check the optics
(Fig. 7-18).
• At the completion of the test, a printout of the
results will be printed and the <Remove SRC>
message will be displayed (Fig. 7-19).