A-2 Operator’s Manual – OPTI CCA-TS
APPENDIX A - TECHNICAL SPECIFICATIONS
Operating Parameters
Minimum Sample Size 125µL (60µL for B60 cassette)
Sample Type heparinized whole blood, plasma or serum
Sample Application syringe, capillary or ComfortSampler
Sample Input automatic aspiration
Analysis Time < 2 minutes, typically approx. 1 minute
to result
Ambient Temperature Range 10 °C - 32 °C (50 °F - 90 °F)
Relative Humidity Range 5% - 95% (non-condensing)
TypeofMeasurement opticaluorescence,
for tHb/SO
2
opticalabsorbance/reectance
Input Values
Patient ID 15 alphanumeric characters
Operator ID 11 alphanumeric characters
Accession Number 12 alphanumeric characters
Patient Temperature 14 – 44° C (58 - 111°F)
Patient Sex male, female or ?
Date of birth MMM-DD-YYYY
Hemoglobin Type adult or fetal
Puncture Site LR/RR/LB/RB/LF/RF/Cord/Scalp, where:
LR = Left Radial
RR = Right Radial
LB = Left Brachial
RB = Right Brachial
LF = Left Femoral
RF = Right Femoral
Cord = Cord
Scalp = Scalp
Bypass Off Pump / On Pump
Sample Type Art/Ven/MixVen/Cap/Cord/CPB, where:
Art = Arterial
Ven = Venous
MixVen = Mixed Venous
Cap = Capillary
Cord = Cord
CPB = Cardio-Pulmonary Bypass
Total Hemoglobin, tHb 1 - 26 g/dL / 10 – 260 mg/dL /
0.6 – 16.1 mmol/L
Mean corpuscular hemoglobin 29.0 – 37.0 %
concentration, MCHC%