3-24 Operator’s Manual – OPTI CCA-TS
Fig. 3-41 Interface
Your OPTI CCA-TS has an RS232 standard serial
interface with a baud rate xed at 9600 and an
Ethernet port. These ports may be congured for
ASCII and ASTM output.
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press the
<Miscellaneous> tab and then <Hardware>.
3. In the <Interface> screen (Fig. 3-41), you can
select <External Barcode Enable>.
4. Select the communication <Format>:
<ASCII>, <ASTM>, or <CF> .
• <ASCII> - Data in easy to read OPTI Medical
custom format. The OPTI CCA-TS exports data
string identical to the internal printer output.
• <ASTM> - Complies with ASTM standard.
Please refer to OPTI CCA-TS interface
specications for more information.
• <CF> - Compact Flash is used for archiving
data to a Compact Flash Card. An Export Kit is
available (BP7140) with a properly formatted
card, instructions and card reader. The database
can be exported to the Compact Flash card in
CSV format and then read by a PC.
Select <Comma> or <Semicolon> for your
CSV le delimiter in MS Excel.
• <Link> - Select <Serial> or <Ethernet>.
5. The <Ethernet> screen (Fig. 3-42) is used to
congure Ethernet settings: <Format>, <IP
Type>, <IP Address>, <Subnet Mask>,
<UDP Port>, <TCP Port>, and <Host IP>.
NOTE: The analyzer must be connected to an active
network to activate this menu.
3.3.3.5 Setting Up Communications
3.3.3.5.1 ConguringtheCommunicationFormat
Fig. 3-42 Ethernet Settings