3-26 Operator’s Manual – OPTI CCA-TS
Fig. 3-47 Select Printer
3.3.3.6 Setting the Printer
Fig. 3-48 Printer Settings
This menu allows you to program the printing
functions of your analyzer.
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press the
<Miscellaneous> tab and then <Printer>
(Fig. 3-47).
On the <Settings> tab (Fig. 3-48), you can select
to have a patient report printed at the end of each
measurement. The second option lets you add a
calibration report to each patient report. The next
option lets you add reference ranges to each patient
report. The last option lets you select how many
copies will be printed.
3. Select the options to be enabled.
4. Press to accept the changes.
5. Press to return to the <Setup> screen
or
to return to <Ready>.
NOTE: This setting affects the patient report only.
All other print functions are still active, even
if the patient report is not activated.
Fig. 3-46 Ready Display
The <Ready> screen displays whether Compact
Flash Card and network connection are enabled
(Fig. 3-46) using the following icons:
- The compact ash card is writable.
- The compact ash card is not writable.
- The network is connected.
- The network is not connected.