Operator’s Manual – OPTI CCA-TS 4-15
4 CALIBRATION AND QUALITY CONTROL
4.5.5 Sending Data to a Computer
The OPTI CCA-TS provides you with the ability to
export Patient and QC information to a connected
computer or HIS/LIS.
Prior to sending data to a computer the OPTI CCA-
TS communication port must be congured (See
section 3.3.3.5) and a physical connection to the
receiving computer must be made.
Before exporting to the Compact Flash (CF) Card
(included in CF Export Kit BP7140), make sure that
the Compact Flash card is properly inserted in the
CF port.
1. In the<Ready> display, select
<DataManager>.
2. Select <Patient>, <SRC> or <Controls>
(Fig. 4-32).
3. Select the data to be exported and press
to start the data transfer (Fig. 4-33).
A warning will be displayed asking you to conrm
your choice (Fig. 4-34).
Fig. 4-32 Select Data
Fig. 4-33 Export Data
Fig. 4-34 Export Selected Data
WARNING - Export Selected Control Data?
WARNING - Export Selected Control Data?