4-14 Operator’s Manual – OPTI CCA-TS
4 CALIBRATION AND QUALITY CONTROL
4.5.4 Printing Control Reports
Your OPTI CCA-TS can print control measurement
reports and statistical information.
1. In the <Ready> display, select
<DataManager>.
2. Press <Controls> (Fig. 4-29).
3. In the<Data-ControlMeasurement>
screen (Fig. 4-30), press the
button to
display the Control results (Fig. 4-31). Use the
and buttons to display the
previous or next page of results.
4. To print individual results, highlight the desired
measurement (Fig. 4-30). To print groups
of results, highlight the rst measurement
to be printed, press
, then select
the last measurement to be printed. All the
measurements in between will be selected.
Press
to select all results.
5. Press to print your selection.
6. Press the <Statistics>button to print
a statistical report of the last 30 control
measurements or more, if available in database.
8. After printout, the database can be deleted by
pressing .
9. Before the database is deleted, enter the
password to initiate the procedure, if a password
has been activated under <Setup>.
10. Press to return to the <Ready> display.
Fig. 4-29 Select Controls
Fig. 4-30 Controls Measurement
Fig. 4-31 View Control Results