Operator’s Manual – OPTI CCA-TS 4-13
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-27 Accept QC Results
Fig. 4-26 QC Results
Upon completion of the measurement, the results are
displayed (Fig. 4-26).
• The OPTI CCA-TS Analyzer will indicate
whether the values are within or outside the
programmed ranges with a <Pass/Fail>
display next to the parameter label.
• Results obtained are applicable to the sensor
cassette type being used for patient sample
(E-Ca Type shown for reference).
16. Press to accept or reject results.
• Press <Accept> if results are acceptable
(Fig. 4-27), and the results will be stored in
the Control Database.
• Select <Reject> to reject the results.
Rejected results will not be stored in the
Control Database.
• Select <Review> to view the results again.
In either case, the results will be printed when
the data input is complete. Please follow the
regulatory guidelines of your hospital for
documenting corrective action, if results are
rejected.
If any of the results are outside of the OPTI
CCA-TS’s measurement range, giving a ‘LOW’
or ‘HIGH’, the results cannot be accepted to the
controls database.
17. Open the sample chamber cover and remove the
cassette (Fig. 4-28).
• If other levels of controls are to be run,
repeat the procedure.
Fig. 4-28 Open SMC Cover and
Remove Cassette
Open Measurement Chamber
Cover and Remove Cassette
QC->Control
Open Measurement Chamber
Cover and Remove Cassette
QC->Control