Operator’s Manual – OPTI CCA-TS 3-3
3 CUSTOMIZATION
3.3.1 QC Setup
3.3.1.1 Setting Up the Standard Reference Cassette (SRC)
When you open a new SRC, the lot number
should be entered into the analyzer, along with the
expiration date. Each SRC level of control has its
own unique lot number printed on the pouch.
NOTE: The procedure for programming SRC QC
ranges as described below is identical for
all levels.
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. Enter the password if this security function has
been activated under <Setup>.
3. In the <System Setup> menu, press <SRC>
(Fig. 3-6).
4. Take an SRC pouch and scan the bar code by
holding it 2-3 inches (5-8 cm) from the bar code
scanner located on the bottom right-hand corner
of the analyzer (Fig. 3-7).
• The red line from the bar code scanner
should cover the entire bar code.
• A beep indicates a valid bar code.
• A red status light indicates an invalid bar
code (e.g. SRC expired).
NOTE: If the bar code is damaged or unreadable,
enter the bar code digits using the keypad.
5. If the level, lot number and expiration date are
correct, press to accept (Fig. 3-8).
Fig. 3-6 Select SRC
Fig.3-8ConrmSRCInformation
Fig. 3-7 Scan Bar code