Operator’s Manual – OPTI CCA-TS 7-29
7 DIAGNOSTICS AND TROUBLESHOOTING
Fig. 7-43 Data Manager
This option allows you to print calibration reports.
From the <Ready> display, press
<Data Manager> to access the <Data> screen
(Fig. 7-43).
The section <Calibration Reports> contains the
following reports:
<Patient>, <SRC>, <Controls> and <Errors>.
7.2.22 Calibration Reports
Fig. 7-44 Select Measurement
The <Patient Calibration Report> contains
information about the measured signal in millivolts
and drifts observed during measurement.
• To print a patient calibration report, select
<Calibration Reports -> Patient> in the
<Data> screen (Fig. 7-43).
• In the <Data - Patient Diagnostics> screen
(Fig. 7-44), select the desired measurement and
press the
button to display the results
(Fig. 7-45). Use the and
buttons to display the previous or next page of
results.
• To print individual results, highlight the desired
measurement (Fig. 7-44). To print groups
of results, highlight the rst measurement
to be printed, press
, then select
the last measurement to be printed. All the
measurements in between will be selected. Press
to select all results.
Press to print your selection.
• Press
to return to the <Data> screen.
7.2.22.1 Patient Calibration Report
Fig. 7-45 Patient Diagnostics