EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - Entering the Barometric Pressure

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS 3-21
3 CUSTOMIZATION
The <Hardware> menu consists of three screens,
<Settings>, <Interface> and <Ethernet>.
The <Settings> screen is used to adjust your local
barometric pressure, the audible alarm, and battery
save mode.
The <Interface> and <Ethernet> screens can be
used to congure communication settings.
To adjust the tracking barometer within the OPTI
CCA-TS, follow the instructions below:
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press the
<Miscellaneous> tab and then <Hardware>
to select this function (Fig. 3-37).
3. In the <Settings> menu, press to enter
an offset from the true barometric pressure
(Fig. 3-38).
4. Type in the new numbers and press
to accept the value.
5. Press to return to the <Setup> screen
or
to return to <Ready>.
CAUTION:
Use the absolute barometric pressure
and not the altitude-corrected pressure
(check with your local weather service
or airport).
NOTE: You may change barometric pressure units
from mmHg to mbar (See section 3.3.2.6).
3.3.3.2 Entering the Barometric Pressure
Fig. 3-37 Select Hardware
Fig. 3-38 Enter barometric pressure

Table of Contents