Operator’s Manual – OPTI CCA-TS 3-21
The <Hardware> menu consists of three screens,
<Settings>, <Interface> and <Ethernet>.
The <Settings> screen is used to adjust your local
barometric pressure, the audible alarm, and battery
save mode.
The <Interface> and <Ethernet> screens can be
used to congure communication settings.
To adjust the tracking barometer within the OPTI
CCA-TS, follow the instructions below:
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press the
<Miscellaneous> tab and then <Hardware>
to select this function (Fig. 3-37).
3. In the <Settings> menu, press to enter
an offset from the true barometric pressure
(Fig. 3-38).
4. Type in the new numbers and press
to accept the value.
5. Press to return to the <Setup> screen
or
to return to <Ready>.
CAUTION:
Use the absolute barometric pressure
and not the altitude-corrected pressure
(check with your local weather service
or airport).
NOTE: You may change barometric pressure units
from mmHg to mbar (See section 3.3.2.6).
3.3.3.2 Entering the Barometric Pressure
Fig. 3-37 Select Hardware
Fig. 3-38 Enter barometric pressure