8-6 Operator’s Manual – OPTI CCA-TS
8 OPERATING PRINCIPLES
8.5.2 Test Conditions
Sample Size: a minimum of 125 µL (60µL for B60 cassette)
Sample Type: heparinized whole blood, serum and plasma
Sample Application: syringe, capillary or ComfortSampler
Ambient Temperature: 10 - 32 ºC (50 – 90 ºF)
Relative Humidity: 5% to 95% (non-condensing)
Type of Measurement: optical uorescence (pH, PO
2
, PCO
2
, Na
+
, K
+
, Ca
++
) Cl
-
,
Glu, BUN (urea), Lac) and reectance (tHb, SO
2
)
8.5.3 Input Values
Parameter Range Default Display Units
Resolution
Patient ID 15 alphanumeric characters Blank
Operator ID 11 alphanumeric characters Blank
Accession Number 12 alphanumeric characters Blank
Patient temperature, T 14 to 44 37.0 0.1 °C
58 to 111 0.1 °F
Patient Sex Male, female or ? ?
Hemoglobin type adult or fetal adult
DOB MMM-DD-YYYY
Puncture Site LR/RR/LB/RB/LF/RF/ LR
Cord/Scalp, where:
LR = Left Radial
RR = Right Radial
LB = Left Brachial
RB = Right Brachial
LF = Left Femoral
RF = Right Femoral
Cord = Cord
Scalp = Scalp
Bypass Off Pump / On Pump Off Pump

Table of Contents