Operator’s Manual – OPTI CCA-TS 8-7
8 OPERATING PRINCIPLES
Parameter Range Default Display Units
Resolution
Sample Type Art, Ven, MixVen, Cap, Art
Cord, CPB, where:
Art = Arterial
Ven = Venous
MixVen = Mixed Venous
Cap = Capillary
Cord = Cord
CPB = Cardio-Pulmonary
Bypass
Total hemoglobin, tHb 1 to 26 15.0 0.1 g/dL
1 to 260 1 mg/dL
1 to 16 0.01 mmol/L
Mean corpuscular hemoglobin 29.0 to 37.0 33.3 0.1 %
concentration, MCHC%
O2 Mode Rm Air, Mask, T-P, NC, Rm Air
Vent, Bag, Hood or Other,
where:
Rm Air = Room Air
Mask = Mask
T-P = T-Piece
NC = Nasal Cannula
Vent = Ventilator
Bag = Bag (manual
resuscitation)
Hood = Hood
Other = Other
FIO
2
0.21 to 1.0 0.21 0.01
respiratory quotient, RQ 0.70 to 2.00 0.84 0.10
P
50
15 to 40 26.7 0.1 mmHg
Vent Mode No, SIMV, PSV, PCV, No
CMV/AC, CPAP, PCIVR,
or BIPAP, where:
No = None
SIMV = Synchronized
Intermittent Mandatory
Ventilation
PSV = Pressure Support
Ventilation