7-18 Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.10 Checking the Factory Settings
This <Fset> function (Fig. 7-11) is designed
exclusively for use by authorized OPTI Medical
personnel.
Fig. 7-11 Factory Settings
7.2.11 Checking the Bar Code Scanner
This option allows you to check the function of the
bar code scanner.
From the <Ready> display, press <System
Manager> and <Diagnostics>.
• Select the <Tests> tab.
Press <Barcode> to start the test (Fig. 7-12).
• To test the bar code scanner, scan a bar code
label of e.g. a sensor cassette (Fig. 7-13).
The display will show a sequence of numbers
(Fig. 7-14). Compare the numbers with those
printed on the cassette bar code label. Matching
information conrms the proper function of the
bar code scanner.
• Press
to return to the <Tests> screen.
• Press
to return to the <System>
screen.
Fig. 7-12 Barcode Test
Fig. 7-13 Scan Barcode
Fig. 7-14 Barcode Test
Scan Barcode
Barcode Test . . .
Scan Barcode
Barcode Test . . .
012345678901234567890123456789
Barcode Test . . .
OK
012345678901234567890123456789
Barcode Test . . .
OK

Table of Contents