Operator’s Manual – OPTI CCA-TS 3-9
3 CUSTOMIZATION
3.3.2.2 tHb Prompt
This menu enables you to enter the tHb value from
another source to the receive calculated parameters
listed below when tHb has been suppressed.
Base excess (BE)
Buffer bases (BB)
Base excess actual (BE
act
)
Standard pH (st.pH)
Standard bicarbonate (st.HCO
3
-
)
Oxygen content (O
2
ct)
1. From the <Ready> display, select
<System Manager> and <Setup>.
2. In the <System Setup> menu, press <Patient
Information Menu> and then <Info 2> tab
(Fig. 3-17).
NOTE: By default the prompt is turned on. When the
prompt is turned off, the analyzer uses a default
value for the following calculated parameters:
BE (Base Excess), BEact (Base excess actual),
st.pH (standard pH) and st. HCO3- (standard
bicarbonate). If tHb is suppressed and the
prompt to enter tHb is turned off then BB and
O2Ct will be suppressed if a tHB value has not
been entered. To enter a tHb value, press the
“Patient Information” button when displayed
and go to page 2 and enter the tHb.
NOTE to B-60 cassette users: BB and O2ct require
a valid tHb value to be accurate. Because tHb
is not measured on the B-60 cassette, these
calculated parameters are not available when
running B-60 cassettes.
Fig. 3-17 Patient info 2
Fig. 3-18 tHb Prompt