EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - 8 Operating Principles; Intended Use; Principles of Procedure

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS 8-1
8 OPERATING PRINCIPLES
8 OPERATING PRINCIPLES
8.1 Intended Use
The OPTI CCA-TS Critical Care Analyzer is intended to be used for the measurement of hydrogen ion
concentration (pH), carbon dioxide partial pressure (PCO
2
), oxygen partial pressure (PO
2
), sodium (Na
+
),
potassium (K
+
), ionized calcium (Ca
++
), chloride (Cl
-
), glucose (Glu), blood urea nitrogen (BUN/urea),
lactate (Lac), total hemoglobin concentration (tHb) and hemoglobin oxygen saturation (SO
2
) in samples
of whole blood, and pH, sodium, potassium, ionized calcium, chloride, glucose and BUN (urea) in serum
and plasma, in either a traditional blood gas, clinical laboratory setting or point-of-care locations by
personnel minimally qualied to perform and report these results.
8.2 Principles of Procedure
Luminescence is the emission of light energy resulting from excited molecules returning to a resting state.
When luminescence is initiated by light, it is commonly referred to as uorescence. When a uorescent
chemical is exposed to light energy of an appropriate color, electrons in the molecules of the uorescent
chemical are excited. A very short time later, the electrons return to a resting state and in this process
sometimes emit a small amount of light energy. This energy is less than the excitation energy and so has a
different color. That is, the emitted light (uorescence emission), is red-shifted from the excitation light,
and is much less intense
.
1
Fluorescent optodes (from optical electrodes) measure the intensity of light emitted from uorescent
dyes exposed to a specic analyte. The emitted light is distinguished from excitation light by means of
optical lters. Because the excitation light energy is kept constant, the small amount of light that results
is changed only by the concentration of the analyte. The concentration of the analyte is determined by
the calculation of the difference in uorescence measured at a known calibration point and that measured
with the unknown concentration of analyte. For a description of the measurement principles of the
individual analytes, please refer to the analyte section of the OPTI CCA-TS Operators Manual.
1
Guilbault GG, Ed., Practical Fluorescence, 2nd Ed., Marcel Dekker, 1990.

Table of Contents