7-22 Operator’s Manual – OPTI CCA-TS
7 DIAGNOSTICS AND TROUBLESHOOTING
7.2.15 Checking the Pump Flow
This option is designed to test the pump cartridge.
From the <Ready> display, press <System
Manager> and <Diagnostics>.
• Select the <Tests> tab.
Press <Flow> to start the test (Fig. 7-23).
• Insert a new cassette (Fig 7-24).
• Close the SMC cover.
• Wait for test results (Fig. 7-25).
• The two numbers indicate the actual ow rates
clockwise and counter clockwise (Fig. 7-26).
If one of the two or both rates are out of range,
the test fails.
• Repeat test or replace the pump cartridge, if
the test fails. See replacement instructions in
Chapter 6.
NOTE: It is possible that the test fails the rst
time, even if the pump cartridge is working
correctly.
• Remove the cassette (Fig. 7-27).
Fig. 7-23 Flow Test
Fig. 7-25 Flow Test in Process
Fig. 7-24 Insert Cassette
Fig. 7-26 Flow Test Fail
Fig. 7-27 Remove Cassette
Insert Cassette
FlowTest . . .
Cancel
Insert Cassette
FlowTest . . .
Cancel
Test in Process . . .
FlowTest . . .
Test in Process . . .
FlowTest . . .
FlowTest . . .
Flow: 4.48 99.9
FAIL: Replace pump
OK
FlowTest . . .
Flow: 4.48 99.9
FAIL: Replace pump
OK
Please Remove the Cassette
FlowTest . . .
Please Remove the Cassette
FlowTest . . .