Operator’s Manual – OPTI CCA-TS 4-5
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-4 Insert SRC
Fig. 4-5 Close Cover
Fig. 4-6 Conrm SRC Information
5. Examine the SRC to ensure it is clean and insert
it into the chamber. Press down to properly seat
the SRC (Fig. 4-4).
6. Close the sample chamber cover (Fig. 4-5).
After the cover has been closed, the instrument will
automatically detect which level of SRC has been
inserted and prompt you to verify the lot number and
level (Fig. 4-6). This information can be found on
both the SRC cassette itself and its storage pouch.
If the information shown on the display is correct,
press
to continue.
If this information is incorrect, press
to
interrupt this sequence and return to
<QCManager>. (See Section 3.3.1.1, “Setting up
the Standard Reference Cassette (SRC)).