Operator’s Manual – OPTI CCA-TS 7-27
7 DIAGNOSTICS AND TROUBLESHOOTING
To set up lactate, you have to run up to 8 ampoules
of OPTI Check level 2 (Fig. 7-38). Please make sure
you have at least 8 ampoules of the same lot of OPTI
Check and 8 lactate cassettes of the same lot on hand
before starting setup.
Press to run the rst sample.
Refer to section 4.5.3.1 for instructions on
running controls.
After each level of OPTI Check is run, you
will return to this screen (Fig. 7-39). You will
not receive a results screen or printout until the
setup procedure is complete. Continue to run the
OPTI check with the same lot of cassettes and
OPTI check until prompted. You may run from 5
to 8 OPTI checks.
After running the required number of samples,
the lactate parameter is enabled (Fig. 7-40).
Press to exit the menu.
You will receive a lactate setup report once
complete.
After completing this procedure, OPTI Medical
recommends that you run two levels of OPTI Check
using B-Lac cassettes to verify performance.
Fig. 7-39 Run Controls
Fig. 7-40 Lactate enabled
Fig. 7-38 Run Controls