pH-B-2 Operator’s Manual – OPTI CCA-TS
pH (Dry Sensor) ANALYTES
Standard Reference Cassette (SRC) Limit Values
LOW NORMAL HIGH Units
7.100 ± 0.02 7.400 ± 0.02 7.600 ± 0.02 pH units
Interferences
Tonometered whole blood samples were spiked with a number of endogenous and exogenous chemicals
and tested for interference following the CLSI guideline EP7-A2
:
Chemical Interferent
Concentration
pH Level Interference
Acetaminophen 1.66 mM
7.170 NO
7.520 NO
Acetylsalicylic acid 3.33 mM
7.170 NO
7.520 NO
Ascorbic acid 0.23 mM
7.170 NO
7.520 NO
B-Hydroxybutyric acid 16.03 mM
7.170 NO
7.520 NO
Bilirubin 0.26 mM
7.170 NO
7.520 NO
Cardiogreen 0.0065 mM
7.170 NO
7.520 NO
Cystein 6.41 mM
7.170 NO
7.520 NO
Ethanol 86.8 mM
7.170 NO
7.520 NO
Evans blue 0.0104 mM
7.170 0.140
7.520 NO
Glycolic acid 10 mM
7.170 NO
7.520 NO
Halothane 0.759 mM
7.170 NO
7.520 NO
Ibuprofen 2.43 mM
7.170 NO
7.520 NO
Intralipid 1%
7.170 NO
7.520 NO
Methylene Blue 0.125 mM
7.170 NO
7.520 0.033
Sodium Chloride 20 mM
7.170 NO
7.520 NO

Table of Contents