EasyManua.ls Logo

OPTIMedical OPTI CCA-TS - Page 53

OPTIMedical OPTI CCA-TS
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual – OPTI CCA-TS 3-13
3 CUSTOMIZATION
7. Press <Limits 3> to advance to the next screen
(Fig. 3-26) with the remaining parameters.
The instrument is preset to the following ranges of
limit values:
pH: 7.2 - 7.6
PCO
2
: 30 - 50 mmHg
PO
2
: 70 - 700 mmHg
Na
+
: 135 - 145 mmol/L
K
+
: 3.5 - 5.1 mmol/L
Cl
-
: 95 - 115 mmol/L
Ca
++
: 1.12 - 1.32 mmol/L
Glu: 60.0 - 120.0 mg/dL
Glu: 3.3 - 6.6 mmol/L
BUN: 5.9 - 19.9 mg/dL
Urea: 2.1 - 7.1 mmol/L
Lac: 0.90 - 1.70 mmol/L
tHb: 12 - 17 g/dL
SO
2
: 90 - 100 %
Units may be changed (See section 3.3.2.6).
In all data input screens, if unreasonable
numbers are entered, the system automatically
ags the error and displays the valid range.
If you wish to turn off limits agging,
enter the system ranges for each parameter.
For instance, for pH, the low is 6.600 and
the high is 7.800 (See Analytes Section for
specications of the reportable ranges for each
parameter measured).
The limits entered here will reside in the
instrument memory even after system power is
turned off.
8. Press to return to the <Setup> screen
or
to return to <Ready>.
Fig. 3-26 Enter Limits 3

Table of Contents