Operator's Manual
Issue 05 /
Feb 2015 / UMC0071
MercuryiTC
©2014 Oxford Instruments Omicron NanoScience. All rights reserved.
Page 50
4 Tap the Sweep button. The Sweep Table page opens.
5 Tap
to add a row to the table.
6 Enter the temperature (in K) for the first step in the FinalT(K) parameter box in row 1 of the
Sweep Table.
7 Enter the time for the first ramp (in minutes) in the Time To Final T(mins) parameter box in
row 1.
8 Enter the time to hold the step temperature (in minutes) in the Hold At Final T(mins)
parameter box in row 1.
9 Tap to add another row to the table and repeat steps 6 – 8.