3. At the Calibration Check screen, select the correct energy test mode (W-hr, Var-hr, or Va-
hr) and enter the desired time and Kh value of the standard. When using a series parallel
load, the Kh value should be multiplied by 3 (ex. 0.00001 * 3 = 0.00003)
4. Press F6 (Start Test) to begin the test.
5. View and record results.
6. Press PREV to return to the Calibration Verification menu.