13.3.1 Calibration Verification Setup
Functionality:
Moves to next field
Moves to previous field
Enters drop down box and views test mode selections; checks/unchecks
check boxes
Moves cursor up and down for selecting a test mode in the drop down box
Continues
Description:
This feature can perform a cross-check against the user’s shop standard to verify the accuracy.
Enter the Kh value of the shop standard into the field for “Standard (Kt)”. Remember if the
standard is single phase and you have the system hooked up series-parallel then you need to
multiply the standard Kt by three (3). The user can also enter the desired test time and energy
test mode (Wh, VAh, and VARh). When selecting the Energy Test Mode, verify the shop
standard is correctly matched to the PowerMaster
®
’s energy test mode.
Using Standard Voltage and Current Cables with the PowerMaster
®
What Do I Need?
1. PowerMaster
®
unit
2. Standard current and voltage leads
Rev 1.5 116