EasyManuals Logo

PP Systems TARGAS-1 User Manual

PP Systems TARGAS-1
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #135 background imageLoading...
Page #135 background image
TARGAS-1 Operation Manual V. 1.02 135 support@ppsystems.com
49
REF_ON_TIME
Time in Ref state
3 - 15
5.0
50
AN_ON_TIME
Time in An state
3 - 15
5.0
51
DEVICE_MODE
The type of device that is
expected to connected
0 = Diff
1 = PLC5
2 = Closed
3 = Absolute
4 = PLC3
5 = SRC??
6 = CPY??
1
52
LEAF_AREA
Area of the leaf exposed
in the cuvette
0.5 4.5
4.5
53
RB
PLC Value
0.20 0.90
0.30
54
RS_FACT
0.10 -1.00
0.5
55
TEMPERATURE_CONTROL
Temperature setting
when the temperature
mode is 1
0 - 45
25
56
TEMPERATURE_MODE
How the temperature in
the chamber is controlled
0= none
1= Chamber
2= track Amb
3= leaf
4= leaf Amb
1
57
PAR_CONTROL
Par setting for Light unit
0-2500
500
58
RED_PERCENT
0-100
25
59
GREEN_PERCENT
0-100
25
60
BLUE_PERCENT
0-100
25
61
WHITE_PERCENT
0-100
25
62
LIGHT_MODE
How the light unit is
controlled
0= no LU
1= Normal
2= not used
3= not used
4= Calibrate
0
63
PAR_SCALE_FACTOR
User Correction Factor of
the Extern PAR (PLC3)
0.8 1.2
1
64
LU_SCALE_FACTOR
User Correction Factor of
the light unit (PLC3)
0.8 1.2
1
65
LEAF_TYPE_MEASUREMENT
What method is used to
determine the
temperature of the leaf
0 = Chamber
1 = IR PLC3
2 = Chamber
3 =Energy
3
66
CO2_SCALE_FACTOR
User Correction Factor of
the CO2 IRGA
0.8 1.2
1
67
H2O_SCALE_FACTOR
User Correction Factor of
the H2O IRGA
0.8 1.2
1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PP Systems TARGAS-1 and is the answer not in the manual?

PP Systems TARGAS-1 Specifications

General IconGeneral
BrandPP Systems
ModelTARGAS-1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals