Remote control commands
R&S
®
NRPxxP
116User Manual 1179.5760.02 ─ 02
SYSTem:RESTart
Restarts the firmware of the power sensor.
Usage: Event
SYSTem:RUTime <update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode, INITiate:CONTinuous ON.
Sets the result update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 0.1
Default unit: Seconds
SYSTem:SERRor?
Queries the next static error, if available.
Static errors, as a rule, prevent the execution of normal measurements.
Errors in SCPI communication are queried using SYSTem:ERRor[:NEXT]?.
Usage:
Query only
SYSTem:SUTime <update_time>
Effective only in the NRP legacy mode. Relevant only in continuous measurement
mode, INITiate:CONTinuousON.
Sets the status update time. That is the maximum rate in which the power sensor can
output measurement results.
Parameters:
<update_time> Range: 0.0 to 10.0
*RST: 10e-3
Default unit: Seconds
SYSTem:TLEVels?
Queries the possible power test levels of the power sensor.
Usage:
Query only
SYSTem:TRANsaction:BEGin
Starts a series of settings.
Configuring the system