Remote control commands
R&S
®
NRPxxP
78User Manual 1179.5760.02 ─ 02
Example:
FREQ 10000
8.7.3 Excluding intervals
In the burst average and timeslot modes, you can define a time interval at the begin-
ning or at the end of an integration interval that is excluded from the measurement
result. See in Figure 8-7.
p
t
2
1
3
4
Figure 8-7: Excluding intervals
1 = Trigger level
2 = Excluding interval at the beginning
3 = Excluding interval at the end
4 = Integration interval
Remote commands:
SENSe<Sensor>:TIMing:EXCLude:STARt..........................................................................78
SENSe<Sensor>:TIMing:EXCLude:STOP.......................................................................... 78
SENSe<Sensor>:TIMing:EXCLude:STARt <exclude_start>
Available in burst average and timeslot modes.
Sets a time interval at the beginning of bursts that is excluded from the measurement.
See Figure 8-7.
Parameters:
<exclude_start> Range: 0.0 to 10.0
*RST: 0.0
Default unit: Seconds
SENSe<Sensor>:TIMing:EXCLude:STOP <exclude_stop>
Available in burst average and timeslot modes.
Configuring basic measurement parameters