Remote control commands
R&S
®
NRPxxP
74User Manual 1179.5760.02 ─ 02
8.7 Configuring basic measurement parameters
The following section describes the settings common for several measurement modes.
8.7.1 Configuring auto averaging
Describes the commands for automatic averaging in continuous average and burst
average measurements.
Remote commands:
SENSe<Sensor>:AVERage:COUNt................................................................................... 74
SENSe<Sensor>:AVERage:COUNt:AUTO..........................................................................74
SENSe<Sensor>:AVERage:COUNt:AUTO:MTIMe...............................................................75
SENSe<Sensor>:AVERage:COUNt:AUTO:NSRatio.............................................................75
SENSe<Sensor>:AVERage:COUNt:AUTO:RESolution........................................................ 75
SENSe<Sensor>:AVERage:COUNt:AUTO:SLOT................................................................ 76
SENSe<Sensor>:AVERage:RESet.................................................................................... 76
SENSe<Sensor>:AVERage:COUNt:AUTO:TYPE................................................................ 76
SENSe<Sensor>:AVERage:TCONtrol................................................................................ 77
SENSe<Sensor>:AVERage[:STATe]...................................................................................77
SENSe<Sensor>:AVERage:COUNt <count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Average count is often also called averaging factor, but it designates the same thing,
i.e the number of measured values that have to be averaged for forming the measure-
ment result.
Averaging is only effective, if SENSe<Sensor>:AVERage[:STATe] ON is set.
Parameters:
<count> Range: 1 to 1048576
*RST: 1024
Example:
AVER:COUN 1
SENSe<Sensor>:AVERage:COUNt:AUTO <state>
Sets how the average count is determined.
Parameters:
<state> ON
Auto averaging: the averaging factor is continuously determined
and set depending on the power level and other parameters.
OFF
Fixed filter: the previous, automatically determined averaging
factor is used.
Configuring basic measurement parameters

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP P Series and is the answer not in the manual?

R&S NRP P Series Specifications

General IconGeneral
BrandR&S
ModelNRP P Series
CategoryAccessories
LanguageEnglish

Related product manuals