Remote control commands
R&S
®
NRPxxP
63User Manual 1179.5760.02 ─ 02
"5 MHZ"
5 MHz
"1.5 MHZ"
1.5 MHz
"300 KHZ"
300 kHz
*RST: "FULL"
SENSe<Sensor>:BWIDth:VIDeo:LIST?
Available in trace and statistics modes.
Lists all available video bandwidth settings.
Usage:
Query only
8.6.4.2 Configuring the trace measurement
SENSe<Sensor>:TRACe:AVERage:COUNt........................................................................63
SENSe<Sensor>:TRACe:AVERage:TCONtrol.....................................................................63
SENSe<Sensor>:TRACe:AVERage[:STATe]....................................................................... 64
SENSe<Sensor>:TRACe:FFT:WINDow..............................................................................64
SENSe<Sensor>:TRACe:MPWidth?.................................................................................. 65
SENSe<Sensor>:TRACe:OFFSet:TIME............................................................................. 65
SENSe<Sensor>:TRACe:POINts.......................................................................................65
SENSe<Sensor>:TRACe:POINts:FPGA?........................................................................... 65
SENSe<Sensor>:TRACe:TIME......................................................................................... 65
SENSe<Sensor>:TRACe:TIME:FPGA?..............................................................................66
SENSe<Sensor>:TRACe:AVERage:COUNt <count>
Available in trace and statistics modes.
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Averaging is only effective, if SENSe<Sensor>:TRACe:AVERage[:STATe] ON is set.
Parameters:
<count> Range: 1 to 65536
*RST: 4
SENSe<Sensor>:TRACe:AVERage:TCONtrol <mode>
Defines how the measurement results are output. This is called termination control.
See also Chapter 8.5, "Controlling the measurement", on page 41.
Parameters:
<mode> MOVing | REPeat
Configuring the measurement modes