Firmware update
R&S
®
NRPxxP
30User Manual 1179.5760.02 ─ 02
During the update process, a progress bar is displayed. The update sequence can
take a couple of minutes, depending on the power sensor model and the size of
the selected file.
8. Check if the update was successful. The firmware version in the "Identification"
field must match the version you selected in the "Firmware" field.
Troubleshooting
You do not find the power sensor in the list of power sensors provided by Firmware
Update for NRP Family.
The driver assigned to the power sensor is the legacy driver.
Install a recent VISA software.
The power sensor is highlighted by a yellow exclamation mark in the Windows device
manager.
Windows tries in vain to find a USB driver for the power sensor.
Install a recent VISA software.
Further information:
Chapter 11.3, "Problems during a firmware update", on page 139
6.2.2 Using remote control
If you want to integrate a firmware update function in an application, use SYSTem:
FWUPdate on page 112.
Updating the firmware