EasyManuals Logo

R&S NRP P Series User Manual

R&S NRP P Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
Firmware update
R&S
®
NRPxxP
31User Manual 1179.5760.02 ─ 02
Example:
You want to update your R&S NRP40P with the NRPxP_02.30.21062301.rsu file.
This file has a size of 10242884 bytes.
To send the file to the power sensor for updating the firmware, your application has to
assemble a memory block containing:
SYST:FWUP <block_data>
The <block_data> is definite length-arbitrary block data as described in SYSTem:
FWUPdate on page 112.
The size of the file is 10242884. This number has 8 digits. Thus, the <block_data>
consist of the following:
#
8
How many digits follow to specify the file size.
10242884
Number that specifies the file size.
<file_contents>
Contents of the *.rsu file, byte-by-byte
0x0a
Delimiter
In this example, you write exactly 10242905 bytes to the power sensor, for example by
using a 'viWrite()' function.
The 10242905 bytes result from the values of the list above:
9 + 1 + 1 + 1 + 8 + 10242884 + 1
In a (pseudo) string notation, the memory block looks as follows:
SYST:FWUP #810242884<file_contents>0x0a,
Updating the firmware

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP P Series and is the answer not in the manual?

R&S NRP P Series Specifications

General IconGeneral
BrandR&S
ModelNRP P Series
CategoryAccessories
LanguageEnglish

Related product manuals