Remote control basics
R&S
®
NRPxxP
132User Manual 1179.5760.02 ─ 02
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
0
Sensor calibration failed
0
0
0
0
0
0
0
0
0
0
0
0
0
0
Figure 10-5: Questionable calibration status register
Querying the register:
STATus:QUEStionable:CALibration:CONDition?
STATus:QUEStionable:CALibration[:SUMMary][:EVENt]?
Table 10-7: Used questionable calibration status bits and their meaning
Bit no. Short description Bit is set if
1 Sensor calibration failed Zeroing of the power sensor was not successful.
10.2.4 Standard event status and enable register (ESR, ESE)
The ESR is already defined in the IEEE 488.2 standard. It is comparable to the EVENt
register of a SCPI register. The standard event status register can be read out by
*ESR?.
The ESE forms the associated ENABle register. It can be set and read by *ESE.