Remote control basics
R&S
®
NRPxxP
131User Manual 1179.5760.02 ─ 02
Table 10-5: Used questionable status bits and their meaning
Bit no. Short description Bit is set if
3 Power summary Summary of Questionable power status register exists.
8 Calibration summary Summary of Questionable calibration status register exists.
9 POST failure (self test) Built-in test of the R&S NRPxxP that is carried out automati-
cally upon power-up has generated an error.
10.2.3.1 Questionable power status register
Contains information whether the measured power values are questionable.
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
+
0
Sensor measurement corrupt
0
0
0
0
0
0
0
0
0
0
0
0
0
0
Figure 10-4: Questionable power status register
Querying the register:
●
STATus:QUEStionable:POWer:CONDition?
●
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
Table 10-6: Used questionable power status bits and their meaning
Bit no. Short description Bit is set if
1 Sensor measurement corrupt Measurement data of the power sensor is corrupt.
10.2.3.2 Questionable calibration status register
Contains information whether the zeroing of the power sensor was successful.
Status reporting system