Remote control commands
R&S
®
NRPxxP
95User Manual 1179.5760.02 ─ 02
32 | 64
32-bit or 64-bit
If you omit the length, the R&S NRPxxP sets the last used
length.
Example for REAL,32 format:
#6768000....<binary float values>....
Example for REAL,64 format:
#71536000....<binary float values>....
<data[,length]> <ASCii[,0 to 12]>
ASCii
List of comma-separated, readable values.
[,0 to 12]
Defines the number of decimal places.
The reset value 0 does not restrict the number of decimal pla-
ces.
Example for ASCii,4 format:
1.2938e-06, -4.7269e-11, ...
*RST: ASCii,0
8.9 Querying measurement results
After the measurement, you can query the measurement results.
8.9.1 Continuous average measurement results
FETCh<Sensor>:ARRay[:POWer][:AVG]?.......................................................................... 95
FETCh<Sensor>[:SCALar][:POWer][:AVG]?........................................................................95
SENSe<Sensor>:[POWer:][AVG:]BUFFer:DATA?................................................................96
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a buffered continuous average measure-
ment.
To configure the measurand, use CALCulate:FEED before the measurement is initi-
ated.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result of the measurand that was configured
before.
To configure the measurand, use CALCulate:FEED before the measurement is initi-
ated.
Querying measurement results