Remote control commands
R&S
®
NRPxxP
96User Manual 1179.5760.02 ─ 02
Usage: Query only
SENSe<Sensor>:[POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast, FETCh<Sensor>[:SCALar][:POWer][:AVG]? returns a result only if
the buffer is full.
Usage: Query only
8.9.2 Burst average measurement results
FETCh<Sensor>[:SCALar][:POWer]:BURSt?
Queries the last valid measurement result.
To configure the measurand, use CALCulate:FEED before the measurement is initi-
ated.
Usage:
Query only
8.9.3 Timeslot measurement results
SENSe<Sensor>:TRACe:DATA? on page 96
FETCh<Sensor>[:SCALar][:POWer]:TSLot?
Queries the last valid measurement result.
To configure the measurand, use CALCulate:FEED before the measurement is initi-
ated.
Usage:
Query only
8.9.4 Trace measurement results
SENSe<Sensor>:TRACe:DATA?....................................................................................... 96
SENSe<Sensor>:TRACe:FFT:DATA?.................................................................................98
SENSe<Sensor>:TRACe:DATA?
Returns the measured trace data in a well-defined format, showing the course of power
over a defined time.
Unlike FETCh<Sensor>[:SCALar][:POWer][:AVG]?, this command takes the set-
tings of SENSe<Sensor>:AUXiliary into account, as explained below.
Command response
Querying measurement results