Remote control commands
R&S
®
NRPxxP
38User Manual 1179.5760.02 ─ 02
Setting parameters:
<number> Range: 0 to 9
*RST: 0
Usage: Setting only
*SRE <register>
Service request enable
Sets the service request enable register to the specified value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<register> Range: 0 to 255
*RST: 0
*STB?
Status byte
Returns the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers a measurement if the following conditions are met:
Power sensor is in the waiting for trigger state.
Trigger source is set to BUS.
See TRIGger:SOURce BUS.
Usage:
Event
*TST?
Selftest
Triggers a selftest of the R&S NRPxxP and outputs the result. 0 indicates that no
errors have occurred.
Usage:
Query only
*WAI
Wait to continue
Prevents the execution of the subsequent commands until all preceding commands
have been executed and all signals have settled.