Remote control commands
R&S
®
NRPxxP
64User Manual 1179.5760.02 ─ 02
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using SENSe<Sensor>:TRACe:
AVERage:COUNt.
*RST:
REPeat
Example:
TRAC:AVER:TCON REP
SENSe<Sensor>:TRACe:AVERage[:STATe] <state>
Enables or disables the averaging filter.
Parameters:
<state> *RST: ON
SENSe<Sensor>:TRACe:FFT:WINDow <window>
Selects a window function for computing a fast Fourier transform (FFT) of the trace.
The resulting power spectrum of the trace is queried by SENSe<Sensor>:TRACe:
FFT:DATA?.
1.0
0.0
-1.0 1.0
4
3
2
1
samples
amplitude
1
= rectangular window
2 = Hamming window
3 = Hanning window
4 = Blackman window
Configuring the measurement modes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP P Series and is the answer not in the manual?

R&S NRP P Series Specifications

General IconGeneral
BrandR&S
ModelNRP P Series
CategoryAccessories
LanguageEnglish

Related product manuals