Remote control commands
R&S
®
NRPxxP
73User Manual 1179.5760.02 ─ 02
Choose a larger width for the level range using SENSe<Sensor>:STATistics:
SCALe:X:RANGe.
Usage: Query only
8.6.6.5 Configuring the scale
Set the level range and its resolution in pixels for the measurement result. For each
pixel, the selected function, CCDF or PDF, is output in W
-1
.
SENSe<Sensor>:STATistics:SCALe:X:POINts.................................................................... 73
SENSe<Sensor>:STATistics:SCALe:X:RANGe....................................................................73
SENSe<Sensor>:STATistics:SCALe:X:RLEVel....................................................................73
SENSe<Sensor>:STATistics:SCALe:X:POINts <points>
Sets the resolution of the measurement result. The defined number of pixels is
assigned to the logarithmic level range, set by SENSe<Sensor>:STATistics:
SCALe:X:RANGe.
Parameters:
<points> Range: Depends on the number of pixels.
*RST: 200.0
SENSe<Sensor>:STATistics:SCALe:X:RANGe <range>
Sets the level range width of the measurement result.
The level range width divided by N–1, where N is the number of pixels, has to equal at
least the highest attainable level resolution, queried by SENSe<Sensor>:
STATistics:SCALe:X:MPWidth?. N is the number of pixels, set by
SENSe<Sensor>:STATistics:SCALe:X:POINts.
Parameters:
<range> Range: Depends on the highest attainable level resolution.
*RST: 50.0
Default unit: dB
SENSe<Sensor>:STATistics:SCALe:X:RLEVel <rlevel>
Sets the lower limit for the level range of the measurement result.
This level can be assigned to the first pixel. The level assigned to the last pixel is equal
to the level of the first pixel plus the level range.
Parameters:
<rlevel> Range: Depends on the level range.
*RST: -30.0
Default unit: dBm
Configuring the measurement modes

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S NRP P Series and is the answer not in the manual?

R&S NRP P Series Specifications

General IconGeneral
BrandR&S
ModelNRP P Series
CategoryAccessories
LanguageEnglish

Related product manuals