Remote control commands
R&S
®
NRPxxP
76User Manual 1179.5760.02 ─ 02
SENSe<Sensor>:AVERage:COUNt:AUTO:SLOT <slot>
Available only in timeslot mode.
Sets a timeslot from which the measured value is used to determine the filter length
automatically. The timeslot number must not exceed the number of the currently set
timeslots.
Parameters:
<slot> Range: 1 to 16
*RST: 1
SENSe<Sensor>:AVERage:RESet
Deletes all previous measurement results that the averaging filter contains and initial-
izes the averaging filter. The filter length gradually increases from 1 to the set averag-
ing factor. Thus, trends in the measurement result become quickly apparent. Note that
the measurement time required for the averaging filter to settle completely remains
unchanged.
Use this command if:
●
High averaging factor is set.
SENSe<Sensor>:AVERage:COUNt
●
Intermediate values are output as measurement results.
SENSe<Sensor>:AVERage:TCONtrol MOVing
●
Power has significantly decreased since the previous measurement, for example
by several powers of 10.
In this situation, previous measurement results, which are still contained in the averag-
ing filter, strongly affect the settling of the display. As a result, the advantage of detect-
ing trends in the measurement result while the measurement is still in progress is lost.
Example:
AVER:RES
Usage: Event
SENSe<Sensor>:AVERage:COUNt:AUTO:TYPE <type>
Sets the automatic averaging filter.
Parameters:
<type> RESolution | NSRatio
RESolution
Usually used.
NSRatio
Predefines the compliance to an exactly defined noise compo-
nent. Enter this value using SENSe<Sensor>:CORRection:
OFFSet.
*RST: RESolution
Configuring basic measurement parameters