EasyManua.ls Logo

R&S NRP P Series - Page 129

R&S NRP P Series
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control basics
R&S
®
NRPxxP
129User Manual 1179.5760.02 ─ 02
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Sum of bits 1 to 4
Measurement not possible
Erroneous result
Warning
Critical
0
0
Legacy locked state
0
0
0
0
0
0
0
0
Figure 10-2: Device status register
Querying the register:
STATus:DEVice:CONDition?
STATus:DEVice[:EVENt]?
Querying the static errors:
SYSTem:SERRor?
Table 10-4: Used device status bits and their meaning
Bit no. Short description Bit is set if
0 Sum of bits 1 to 4 Sum/combination of static error (SERR) bits 1 to 4.
1 Measurement not possible
2 Erroneous results Static error exists. The measurement result is possibly incor-
rect.
3 Warning Static error exists. Status LED of the power sensor is slowly
flashing red.
Status reporting system

Table of Contents