117Routine operation
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
WARNING!
Incorrect interpretation of test result when using
anticoagulant tubes
To avoid harmful outcomes resulting from the incorrect
interpretation of a test result.
r When using anticoagulant tubes, do not store the
tube after drawing, but apply the blood immediately.
u Entering a patient (113)
u Performing a QC test (122)
u About performing a glucose test (52)
d
m Glucose test strips
m Freshly obtained blood sample
m New pair of disposable gloves
m Waste container
j
m You are logged on to the instrument
m Specified QC tests have been completed successfully
u Logging on to the Glucose app (99)
r To perform a glucose test
Tests
Glucose
1 In the Tests screen, tap the Glucose button.
2 Enter the patient ID using one of the following
methods:
RFID tag
Barcode scan
Keyboard entry
5 Routine operation