135Non-routine operation
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Performing a linearity test
Linearity tests help you check the function and accuracy
of the entire system over the full range of specified
values.
A linearity test comprises the following main steps:
Checking the lot number of the linearity material.
Checking the lot number of the test strips.
Perform the test with a minimum of 3 linearity levels.
i
When performing a linearity test, the instrument
does not perform a linear regression.
c
Determined by your healthcare facility.
d
m Linearity kit
m Glucose test strips
m New pair of disposable gloves
m Waste container
j
m The instrument is turned on.
m You have logged on to the instrument.
r To perform a linearity test
Tests
Linearity
1 In the Tests screen, tap the Linearity button.
2 In the Linearity test screen do one of the following:
Scan the linearity kit lot barcode.
If enabled, tap the tab and then below
Choose linearity kit lot, tap the lot you will use
to perform the test.
6 Non-routine operation