20 Warning messages
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Warning messages
List of warning messages
Failure to observe warning messages may result in
death or serious injury.
r Before using the system, read the warning messages
carefully.
Incorrect test measurement
To avoid harmful outcomes resulting from incorrect test
measurements, ensure the following:
r Only use a blood sample when performing a test
measuring blood glucose.
r Only use an appropriate QC material when
performing QC tests.
r Always wear a new, clean pair of disposable gloves
for each test you perform.
r Use only parts and devices approved by Roche
Diagnostics.
r Do not use the instrument outside the specified
temperature range.
r Do not use the instrument in a hyperbaric chamber.
r Ensure that there are no traces of food on the fingers
or fatty residues from hand cremes or soap products
that may contaminate the sample.
r Ensure that there are no residues of water or
disinfectant on the skin that can dilute the drop of
blood.
r Test strips are single use only. Never reuse a test
strip.
1 General safety information