177Maintenance
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Replacing the battery
When the battery no longer functions correctly it should
be replaced.
When shipped, the battery is installed in the instrument.
Unused batteries lose their charge over time and must be
recharged before they can be used. After installing a new
battery, the instrument should be charged for at least 8
hours (e.g., overnight) in the charging station before
testing.
Replace the battery within approximately 10 minutes to
retain the date and time settings. Beyond this period of
time, and depending on the instrument’s configuration:
• You must re-enter date and time.
• The date and time is set automatically by the DMS.
NOTICE!
Disposal of used batteries
r Do not dispose of the batteries with normal domestic
waste.
r Dispose of used batteries in accordance with
applicable local regulations and directives and your
healthcare facility’s guidelines on the disposal of
electronic waste equipment.
r As a component of the cobas
®
pulse system, the
battery must be deemed as potentially biohazardous.
Treat it accordingly.
c
As necessary
n
< 10 minutes
d
m New battery (only official replacement battery
supplied Roche)
j
m Temperature range of 3-42 °C or 37-108 °F
m Instrument is turned off.
7 Maintenance