87Specifications
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Environmental conditions
The location must comply with the following conditions:
International (Europe) US / Canada
Ambient temperature
during operation
Instrument 12-40 °C 54-104 °F
Charging station 12-40 °C 54-104 °F
Power supply 0-45 °C 32-113 °F
Carry case 12-40 °C 54-104 °F
Ambient temperature
during long-term storage
Instrument 5-40 °C 41-104 °F
Charging station 5-40 °C 41-104 °F
Power supply -40-70 °C -40-158 °F
Carry case 5-40 °C 41-104 °F
Ambient Temperature short
term transport
Instrument -25-70 °C -13-158 °F
Charging station -25-70 °C -13-158 °F
Power supply -40-70 °C -40-158 °F
Carry case -25-70 °C -13-158 °F
Ambient humidity during
operation
Instrument 10-90% (non-condensing)
Charging station 10-90% (non-condensing)
Power supply 0-95% (non-condensing)
Carry case 10-90% (non-condensing)
Ambient humidity during
long-term storage
Instrument 5-85% (non-condensing)
Charging station 5-85% (non-condensing)
Power supply 5-95% (non-condensing)
Carry case 5-85% (non-condensing)
Maximum operating height
(above sea level)
Instrument 4,300 m 14,107 feet
Charging station 4,300 m 14,107 feet
Power supply 4,300 m 14,107 feet
Sample temperature
Instrument Refer to the package insert for information
y Environmental conditions
3 Specifications