176 Cleaning and disinfecting the carry case
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Cleaning and disinfecting the carry case
Cleaning and disinfecting the carry case thoroughly
ensures that nothing can be transmitted to its contents.
After cleaning and disinfecting the carry case, ensure
that it has dried thoroughly before using it again.
c
As specified by your healthcare facility.
d
m Disposable gloves
m Soft, lint-free cloth or gauze
m Pre-moistened wipes for cleaning and disinfection
m Cleaning solutions as per the manufacturer’s product
labelling
j
m The carry case is completely empty
u Acceptable cleaning solutions and disinfectants
(171)
r To clean and disinfect the carry case
1 Wipe the carry case surface with a soft cloth slightly
dampened (not wet).
I
If spraying the carry case, place it on a flat surface
or table.
2 CAUTION! Do not place the instrument in the carry
case until all cleaning/disinfecting steps are
completed.
Allow the carry case to air dry for the recommended
contact time according to the cleaning/disinfecting
solution product labeling.
3 Dry the carry case thoroughly with a dry cloth or
gauze. Visually verify that no solution is in the carry
case after cleaning/disinfecting.
7 Maintenance