138 Performing a linearity test
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
Linearity test
Level 1
Level 2
Level 3
Level 4
Level 5
Level 6
Passed
Required
Required
Required
Required
Required
9 Tap the Proceed button.
f The Linearity test screen is displayed, indicating
if level 1 has passed or failed.
10 To test the remaining 5 levels, do the following:
Tap the next required level button (for example,
the Level 2 button).
Perform runs 1 and 2 for leach remaining level by
repeating steps 4 to 9.
11 After performing run 2 of level 6, you are prompted to
eject the test strip.
f The Linearity test > Result screen is displayed.
Linearity test
Glucose
Result
All levels passed
Confirm
No comments
PASS
Linearity kit lot number
Test strip lot number
12 You can do the following:
Tap the button to view the results of all runs
for all levels in the Result details dialog box.
Tap the button to add comments to the result.
13 Tap the Confirm button to close the Linearity test
screen and return to the Tests screen.
I
You can view the test result at any time in Local
results > QC results.
u Related topics
Reviewing QC test results (126)
6 Non-routine operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas pulse and is the answer not in the manual?

Roche cobas pulse Specifications

General IconGeneral
BrandRoche
Modelcobas pulse
CategoryMedical Equipment
LanguageEnglish

Related product manuals