157Non-routine operation
Roche Diagnostics
cobas
®
pulse · Software version 01.03 · User Assistance · Publication version 1.0
The instrument can be configured to perform a QC test
the first time a test strip from a new lot is inserted into
the instrument.
CAUTION!
Instrument is not connected to cobas
®
infinity edge
If the instrument is not connected to
cobas
®
infinity edge you cannot add a test strip lot as
described in this task.
r You must scan the QR code generated in the
cobas
®
infinity edge portal using the >
Configuration by barcode option.
The new strip lot is then added to the Lot
management screen.
c
As required
d
m Test strip container and test strips from the new lot
j
m The new lot has been uploaded to the DMS
m Connection to cobas
®
infinity edge
r To add a test strip lot to the
instrument
1 In the Glucose app, tap > Lot management.
Lot management
QC lotsTest strip lots
Lot No. Type Expiry date
Add test strip lot
GLU
GLU
GLU
GLU
GLU
2 In the Lot management screen on the Test strip
lots tab, tap the Add test strip lot button.
3 Using the Scan bottle barcode screen, scan the
barcode of the new test strip lot container.
f A message confirms that the new lot has been
added. The new lot is added to the Test strip lots
tab.
4 Do one of the following:
If you are not prompted to perform a QC test, you
can start using the test strips from the new lot.
If prompted, perform a QC test using the new test
strip lot.
After performing the QC test successfully you can
start using the test strips from the new lot.
6 Non-routine operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas pulse and is the answer not in the manual?

Roche cobas pulse Specifications

General IconGeneral
BrandRoche
Modelcobas pulse
CategoryMedical Equipment
LanguageEnglish

Related product manuals